Correlation between the amount of protein analyzed and the accuracy of MW determination using DAWN® DSP and OPTILAB DSP detectors.
Look at the actual data in the "micro analyses" graph
Protein |
[micrograms] |
|
[kDa] a |
[kDa] |
experimental accuracy [%]b |
[kDa] |
Ovalbumin | 150 | 4 | 42.8 | 42.4 | 0.2 to 1.6 | 0.3 |
100 | 7 | 42.8 | 42.3 | 0.2 to 2.4 | 0.8 | |
45-50 | 4 | 42.8 | 41.6 | 0.5 to 5.8 | 1 | |
6-10 | 5 | 42.8 | 42.9 | 1.4 to 4.5 | 1.5 | |
Transferrin | 100 | 3 | 75.2 | 76.5 | 0.7 to 3.2 | 1 |
8 | 5 | 75.2 | 76.3 | 0.3 to 5.2 | 2 |