Search Keck Sites:


W.M. Keck Facility
 Yale University
 300 George Street
 Addresses
 
 Contact Us

Yale University School of Medicine

Keck Home Page > Protein Chemistry > Electrospray Ionization

Mass Spectrometry: Electrospray Ionization

Electrospray ionization mass spectrometry (ESMS) is carried out on a Micromass Q-Tof mass spectrometer on samples with molecular weights from 100 to well over 100,000 (More Information). The technique is sensitive in that it usually can be carried out on low pmol amounts of sample. Samples should be free of detergents, polyethylene glycol type compounds, and salts (e.g., as little as 5 mM NaCl precludes analysis). Upon request, samples that contain salts or other non-volatile reagents may be subjected to a reverse-phase desalting on a ZipTip®, however, this procedure may (particularly in the case of proteins) lead to significant loss of sample. Volatile buffers (such as 1-10 mM ammonium carbonate or acetate) are acceptable and, if necessary, samples may be injected that contain up to 10% (v/v) glycerol. Many different volatile solvent systems can be used (e.g., water or water/organic combinations utilizing acetonitrile, isopropanol, or chloroform/methanol). Unless requested otherwise, all samples will be introduced as a solution in 50% acetonitrile, 0.2% formic acid and will be run with positive ion detection. Since the analysis will fail if the sample is not soluble in this buffer, it is important that the solubility of the sample in 50% acetonitrile, 0.2% formic acid be confirmed prior to sample submission. Typically, 10 µl of a 1-10 µM protein solution is needed for an analysis. The following provides a brief description of several types of commonly requested analyses. In addition, many other types of samples (e.g., carbohydrates, lipids, and polymers) may also be analyzed on the Q-Tof. It is important to note also that the average mass accuracies given below are estimated and that the mass error on any given sample may be either higher or lower than these estimates. Monoisotopic masses should be used for molecules with mass less than 3 kDa while average masses should be used for larger molecules.

Nominal Mass Determinations on Synthetic Molecules Above 150 Da

This type of analysis is extremely useful for rapid monitoring of chemical syntheses where an average mass error of ± 0.2 amu (400 ppm or ± 0.04% for 500 Da) is sufficient to determine if the synthesis is indeed proceeding as planned. This level of mass accuracy can be achieved routinely on the Q-Tof with external calibration and without having to constantly monitor instrument calibration.

Exact Mass Determinations on Synthetic Chemical Intermediates (150-1000 Da)

To reach the extremely high level of mass accuracy required to confirm the elemental composition of the products of chemical syntheses requires the use of internal calibrants. To avoid signal suppression, the concentration of these internal calibrants have to be carefully matched to that of the sample. The average mass accuracies that may be achieved with this approach are ± 0.002 Da over the mass range 150-400 and 5 ppm (± 0.0005%) over the mass range extending from 400-1000)

High Mass Accuracy Determinations on Peptides (100-2500 Da)

Frequent monitoring of the external calibration of the Q-Tof allows the average mass error with external calibration to be decreased to less than ±0.05 amu (50 ppm or ±0.005% for 1000 Da).

ESMS Analysis of Oligonucleotides

ESMS spectra of oligonucleotides are recorded in negative ion mode. To increase sensitivity and mass accuracy the samples should be free of sodium and potassium counterions. There are many published methods to remove or minimize the effect of these counterions. These include organic precipitation, RP-HPLC, ZipTip clean-up (which also rellies on a reverse phase support), or carrying out the ESMS in the presence of triethylamine acetate, piperidine, or imidazole buffers. We use the latter two methods: ZipTip clean-up and/or one of the three buffers mentioned. The longest oligonucleotide we have analyzed was a 39-mer. ESMS spectra of longer oligos have been reported in the literature. We request that a minimum of a 100 pmol be submitted. Any unused sample can be returned if requested beforehand.

ESMS Analysis of Proteins

For large molecular weight biomolecules (e.g., proteins), it is important to recognize that the measured mass is the average mass and that the peak envelope extends over many individual masses. For example, a protein with a mass of 10 kDa. will have a peak envelope that is approximately 20 mass units wide (counting all isotope containing peaks with intensities greater than 1% of the most abundant peak, for a detailed discussion see Anal. Chem. 55,353-356 (1983)). As the size of the molecule increases, the peak envelope gets wider such that the envelope for a 100 kDa protein should contain 57 molecular ion peaks. In instruments that cannot resolve these individual isotopic peaks, the width of this isotopic envelope will determine what mass difference is needed between two molecules in order to detect each molecule. As a general guide, one should be able to theoretically detect two components present in equal amounts if their molecular weight divided by their mass difference is less than approximately 1,000. Hence, if the samples are very clean (i.e., no counterions or adducts which would create 'tails' on the mass spectral peaks) it may be possible to detect a 50,000 Da species in the presence of an equivalent concentration of a 50,050 dalton species.

ESMS spectra are almost always recorded in positive ion mode. The mass spectrum from a protein contains a series of peaks produced from the protein by adding a successive number of protons. Because mass spectrometers record mass-to-charge ratio, electrospray analyses on proteins typically produce a spectrum in the mass range of 500-2000. This spectrum containing multiply charged peaks is processed to produce a peak on a molecular mass scale by either transforming manually identified multiply charged peaks (adjacent peaks can only differ by one charge) or by MaxEnt transformation which only requires an accurate peak width as input. The MaxEnt transformation produces a mock spectrum which can be compared to the raw data to determine the quality of the transformation. The average, expected mass error is about ± 0.01%. (More Information)

 

    Top of Page
Medical Center Yale-New Haven Hospital Yale University

Copyright © 2003, Yale University, New Haven, Connecticut, USA. All rights reserved.
Comments or suggestions to site editor.

Last modified: 23-Oct-2006 (GB)