Analytical Reverse Phase HPLC

Reverse phase HPLC is carried out on a YMC C-18 column (4 mm x 50 mm, 3 micron particle size, 120 angstrom pore size support) that is eluted at 1 ml/min with the following gradient where buffer A is 0.05% TFA in water and buffer B is 80% acetonitrile in 0.05% TFA in water. Peptides are detected by their absorbance at 210 nm.

Time (min)

%Buffer A

%Buffer B

0

98

2

63

62.5

37.5

95

25

75

105

2

98

112

2

98

119

98

2

250

98

2

 

    Top of Page