Keck
Home Page >
Small Scale
Peptide Synthesis > HPLC
Analytical Reverse Phase HPLC
Reverse phase HPLC is carried out on a YMC C-18 column (4 mm x 50 mm, 3 micron
particle size, 120 angstrom pore size support) that is eluted at 1 ml/min with the
following gradient where buffer A is 0.05% TFA in water and buffer B is 80% acetonitrile
in 0.05% TFA in water. Peptides are detected by their absorbance at 210 nm.
|
Time (min) |
%Buffer A |
%Buffer B |
0 |
98 |
2 |
63 |
62.5 |
37.5 |
95 |
25 |
75 |
105 |
2 |
98 |
112 |
2 |
98 |
119 |
98 |
2 |
250 |
98 |
2 |
|